E4990A Help
580
Preset Value 1
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Long
Var= 1
Ana.WriteString ":SERV:CHAN1:TRAC1:LIM:DATA?", True
Var=Ana.ReadNumber
Equivalent Softkey
No equivalent key is available on the front panel.
COM Command Syntax (Internal Control Only)
SCPI.SERVICE.CHANNEL(Ch).TRACE(Tr).LIMIT.DATA
:SERV:CHAN<Ch>:TRAC<Tr>:MEMV
Type
Command/Query
Syntax
:SERVice:CHANnel<Ch>:TRACe<Tr>:MEMValid {ON|OFF|1|0}
:SERVice:CHANnel<Ch>:TRACe<Tr>:MEMValid?
Description
This command sets/gets the memory trace data valid flag. Always set to 0
after preset.
Variable
Parameter Selection Option
Description Memory trace data
Data Type Boolean type (Boolean)
Range ON|OFF|1|0
Preset Value OFF
Query Response

Table of Contents

Other manuals for Keysight E4990A