Programming
581
{1|0}<newline><^END>
Examples
Dim Var as Integer
Ana.WriteString ":SERV:CHAN1:TRAC1:MEMV OFF", True
Ana.WriteString ":SERV:CHAN1:TRAC1:MEMV?", True
Var=Ana.ReadNumber
Equivalent Softkey
No equivalent key is available on the front panel.
COM Command Syntax (Internal Control Only)
SCPI.SERVICE.CHANNEL(Ch).TRACE(Tr).MEMVALID
:SERV:LOGG:CLE
Type
Command
Syntax
:SERVice:LOGGing:CLEar
Description
This command clear all log files about Event, Power on test, Mech sw,
Overload, FW close, Recovery. (no action if simulator mode)
Examples
Ana.WriteString ":SERV:LOGG:CLE", True
Equivalent Softkey
No equivalent key is available on the front panel.
COM Command Syntax (Internal Control Only)
SCPI.SERVICE.LOGGING.CLEAR
:SERV:SREV
Type
Query
Syntax
:SERVice:SREVision?
Description
This command gets system/spec revision.
Variable
Parameter Long
Description 0: applying current system specifications
Data Type Long integer type (Long)