E4990A Help
586
Query Response
{1|0}<newline><^END>
Examples
Dim Var as Integer
Ana.WriteString ":SOUR1:ALC ON", True
Ana.WriteString ":SOUR1:ALC?", True
Var=Ana.ReadNumber
Related Commands
:SOUR:ALC:COUN
:SOUR:ALC:TOL
Equivalent Softkey
Sweep Setup > OSC Level > ALC > ALC
COM Command Syntax (Internal Control Only)
SCPI.SOURCE(Ch).ALC.STATE
:SOUR<Ch>:ALC:COUN
Type
Command/Query
Syntax
:SOURce<Ch>:ALC:COUNt <Value>
:SOURce<Ch>:ALC:COUNt?
Description
This command sets/gets the AC ALC maximum iteration count.
Variable
Parameter Value
Description Maximum iteration count
Data Type Numeric type (Integer)
Range 2 ~ 10
Preset Value
10
Unit -
Resolution -
Query Response

Table of Contents