E4990A Help
602
This command sets/gets the bias limit minimum. This setting is applied for
all channels and traces.
Variable
Parameter Value
Description Minimum limit
Data Type Numeric type (Real)
Range -40 ~ 40
Preset Value
-40
Unit V
Resolution 1m
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Double
Var= -20
Ana.WriteString ":SOUR:BIAS:VOLT:LIM:MIN " & Str(Var), True
Ana.WriteString ":SOUR:BIAS:VOLT:LIM:MIN?", True
Var=Ana.ReadNumber
Related Commands
:SOUR:BIAS:VOLT:LIM:MAX
Equivalent Softkey
Sweep Setup > DC Bias > Min Limit Voltage
COM Command Syntax (Internal Control Only)
SCPI.SOURCE.BIAS.VOLTAGE.LIMIT.MINIMUM
:SOUR<Ch>:BIAS:VOLT:SPAN
Type
Command/Query
Syntax
:SOURce<Ch>:BIAS:VOLTage:SPAN <Value>
:SOURce<Ch>:BIAS:VOLTage:SPAN?
Description

Table of Contents