Programming
601
This command sets/gets the bias limit maximum. This setting is applied
for all channels and traces.
Variable
Parameter Value
Description Maximum limit
Data Type Numeric type (Real)
Range -40 ~ 40
Preset Value
40
Unit V
Resolution 1m
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Double
Var= 30
Ana.WriteString ":SOUR:BIAS:VOLT:LIM:MAX " & Str(Var), True
Ana.WriteString ":SOUR:BIAS:VOLT:LIM:MAX?", True
Var=Ana.ReadNumber
Related Commands
:SOUR:BIAS:VOLT:LIM:MIN
Equivalent Softkey
Sweep Setup > DC Bias > Max Limit Voltage
COM Command Syntax (Internal Control Only)
SCPI.SOURCE.BIAS.VOLTAGE.LIMIT.MAXIMUM
:SOUR<Ch>:BIAS:VOLT:LIM:MIN
Type
Command/Query
Syntax
:SOURce<Ch>:BIAS:VOLTage:LIMit:MINimum <Value>
:SOURce<Ch>:BIAS:VOLTage:LIMit:MINimum?
Description