E4990A Help
650
Ana.WriteString ":TRIG:AVER?", True
Var=Ana.ReadNumber
Related Commands
:TRIG
Equivalent Softkey
Average > Avg Trigger
COM Command Syntax (Internal Control Only)
SCPI.TRIGGER.SEQUENCE.AVERAGE
:TRIG
Type
Command
Syntax
:TRIGger[:SEQuence][:IMMediate]
Description
This command generates a trigger immediately and executes a
measurement, regardless of the setting of the trigger mode.
This command is different from :TRIG:SING
as the execution of the object
finishes at the time of a trigger.
If you execute this object when the trigger
system is not in the trigger wait state (trigger event
detection state), an error occurs when executed and the
object is ignored.
Examples
Ana.WriteString ":TRIG", True
Related Commands
:TRIG:SOUR
:TRIG:SING
*TRG
Equivalent Softkey
Trigger > Trigger (This is for manual trigger.)
COM Command Syntax (Internal Control Only)
SCPI.TRIGGER.SEQUENCE.IMMEDIATE
:TRIG:POIN
Type
Command/Query
Syntax
:TRIGger[:SEQuence]:POINt {ON|OFF|1|0}
:TRIGger[:SEQuence]:POINt?