Keysight N9912-90001 User’s Guide 7- 11
Calibration for NA, CAT, and VVM Modes
Calibration Method Summary
Calibration Method Summary
O,S,L is ALWAYS the most accurate Cal method. The quality of a O,S,L is
completely dependent on the quality of the OPEN, SHORT, LOAD standards.
Use the correct high quality standards to ensure the most accurate
calibration.
QuickCal is ideal for DTF measurements. QuickCal should be performed
when good accuracy is required and when using a jumper cable or adapter.
QuickCal is NOT recommended when the DUT has 3.5 mm-female or
SMA-female connectors.
Even with the optional Load, phase accuracy begins to degrade when
the return loss is greater than about 20dB.
Present Cal is ONLY useful for DUT connections made at the RF Out port
and most accurate at room temperature.
For highest accuracy, a new QuickCal or O,S,L Cal should be performed:
When the temperature changes more than about 10°F (5°C)
When the connection to the DUT requires a different jumper cable or
adapters.