8- 38 Keysight N9912-90001 User’s Guide
SA (Spectrum Analyzer) Mode
Channel Measurements
8-
Trace State is set to Average.
Detector is set to Average.
Count is set to 15.
Any of these settings can be changed manually during a Channel
measurement. Learn more about these settings starting in Average Type” on
page 8-29.
To change Averaging:
—Press Meas Setup 4
.
—Then Average Enable ON OFF
OFF effectively sets the Average Count to 1.
—Then Average Count
.
Enter a value between 1 and 10,000 using the numeric keypad,
arrows, or the rotary knob.
—Press Enter
.
Traces
Only one measurement trace can be displayed in Channel Measurements.
Channel Power (CHP)
Channel Power measures total power over the specified Integrated BW. The
Integration Bandwidth (IBW) can be adjusted to measure the power over
multiple channels.