17- 4 Keysight N9912-90001 User’s Guide
File Management
Saving and Recalling Files
17-
In the graphic above, the current filename is File_0. The cursor is on the
custom DTF_VSWR keyword. The file will be saved to USB (flash drive). File
Type is “Trace + State” (Instrument State).
How to move the cursor
Use the rotary knob to move cursor on the same row. The cursor will change
rows when it comes to a row end.
Use the arrows to move the cursor up and down the rows.
When the current filename is complete
—Press Done to save the file to the specified filename. If the filename already
exists, a warning is displayed.
—Press Cancel
to close the labeler without saving a file.
If the filename already exists, you are prompted to choose from the
following:
—Press Yes
to overwrite the existing file.
—Press Cancel
to return to edit the filename.
Recall Files
The FieldFox can recall State files, Trace + State files, and image files.
State Files
ALL FieldFox instrument settings for ALL Modes are saved and recalled in a
*.sta file. If error correction is ON in CAT or NA Mode, then calibration data is
also saved and recalled.
If error correction is OFF (CalRdy is active), calibration data is NOT saved in the
*.sta file. When a Trace + State file is recalled on a FieldFox other than the one
it was saved on, differences in the CalRdy calibration data may result in
differences between the displayed traces.
Image Files
The FieldFox can open and display *.png, *.jpg, and *.jpeg images.
How to Recall a File
—Press Save/Recall 9.
—Then Recall
.
Before recalling a file, first set File Type to State or Trace + State, and also
Select Device from which the file will be recalled.