5- 4 Keysight N9912-90001 User’s Guide
NA (Network Analyzer) Mode
NA Mode Settings
5-
Then choose from the following:
Off
(default) No conversion is performed.
Z Conv (--) or Y Conv (--)
Perform conversion for impedance (Z) or
admittance (Y). Then choose from the following for either:
——Auto
The displayed S-parameter is converted to the
appropriate Z or Y parameter: Refl for S11 and S22; Trans for
S21 and S12. When the S-parameter is changed, the
appropriate conversion changes automatically.
——Refl
The displayed S-parameter is converted to Z or Y
reflection, regardless of whether the S-parameter is reflection
(S11 or S22) or transmission (S21 or S12).
——Trans
The displayed S-parameter is converted to Z or Y
transmission, regardless of whether the S-parameter is
reflection (S11 or S22) or transmission (S21 or S12).
1/S
The displayed S-parameter is converted to 1/S-parameter.
Multi-Trace Configurations
In NA Mode you can display multiple traces on the FieldFox screen.
The graphic above shows a x3H configuration. Tr2 is the ACTIVE trace as
indicated by the highlighted Tr2.
Trace Setting Notes
The Frequency Range, IF BW, Resolution, Average, and Output Power
settings are common for all displayed traces.
All other trace settings, such as measurement, format, and limit lines, are
applied individually to the ACTIVE trace in the same manner as when a
single trace is present.