8- 30 Keysight N9912-90001 User’s Guide
SA (Spectrum Analyzer) Mode
SA Mode Settings
8-
How to set Average Count
—Press Meas Setup 4.
—Then Average Count
.
Enter a value from 1 to 10,000 using the numeric keypad, the arrows,
or the rotary knob.
—Press Enter
.
Detection Method
In SA Mode, the X-axis is comprised of data points, also known as “buckets”.
The number of data points is specified using the Points setting. Learn more in
“Points” on page 8-27.
Regardless of how many data points are across the X-axis, each data point
must represent what has occurred over some frequency range and time
interval.
From the frequency span of the measurement, the span of each data point is
calculated as (frequency span / (data points-1)). The detection method allows
you to choose how the measurements in each bucket are displayed.
Figure 8-6 Detection using One Bucket
In the graphic above there is one bucket showing Positive peak, Sample, and
Negative peak Detection methods.