Keysight N9912-90001 User’s Guide 13- 13
Pulse Measurements Mode – Option 330
Pulse Measurement Settings
Auto Analysis
Auto Analysis reports the following TEN pulse measurements:
How to enable Auto Analysis
—Press Mrk ->/Tools.
—Then Auto Analysis
.
Then choose from the following:
OFF
Analysis OFF
ON
The Analysis values appear below trace 1.
1 - Average Power Average power level of all
measured data points
2 - Peak Power Highest reading of all measured
data points
3 - Peak minus
Average
Difference between the above two
results
4 - Pulse Top Power value based on the specified
Pulse Top setting.
5 - Rise Time The time it takes for a pulse to rise
from 10% to 90% of its peak value.
6 - Fall Time The time it takes for a pulse to fall
from 90% to 10% of its peak value.
7 - Pulse Width The time between the rising edge
and falling edge of a pulse.
8 - Pulse Period The time of complete pulse cycle.
9 - Duty Cycle The ratio of the pulse width / pulse
period.
10 - Pulse Rep Freq The rate, in pulses / sec, at which
pulses recur.