Keysight N9912-90001 User’s Guide 5- 15
NA (Network Analyzer) Mode
NA Mode Settings
High (Default setting) Sets output power to the maximum achievable
power at all displayed frequencies. Please see the FieldFox
Specifications (page B-3) for expected power levels.
Low Sets output power to approximately –27 dBm.
Man Set output power to a arbitrary value.
—Then press Nominal Power
.
Then enter a value using the numeric keypad, the arrows, or the
rotary knob.
—Press Enter
.
System Impedance (Z0)
To accurately view data presented in Smith Chart format, first set the System
Impedance.
Learn how to select Smith Chart format in “Format” on page 5-7.
Learn how to make 75 measurements at the FieldFox Supplemental Online
Help: http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
How to set System Impedance
—Press Meas Setup 4
.
—Then Settings
.
Then scroll to System Z0 and press Edit
.
Then type either 50 or 75 and press Enter
.
Nominal here means: a general, descritive term or design parameter. It is
not tested and not covered by the product warranty.
The Nominal Output displayed on the bottom right side of the display
indicates a range of low and high power values. Your actual measured
output power could be a little lower or a little higher.
If you have the USB power sensor feature (Option 302) you can measure
the actual output power for your measurement. Refer to “Power Meter
Mode (Option 302)” on page 11-1.