Keysight N9912-90001 User’s Guide 5- 9
NA (Network Analyzer) Mode
NA Mode Settings
Then choose from the following methods:
Autoscale
Automatically adjusts the Y-axis to comfortably fit the Min
and Max amplitude of the trace on the screen.
Autoscale All
Autoscales all of the traces on the screen, useful only for
multi-trace configurations.
—3.Set Scale, Reference Level, and Reference Position
——Scale
Manually enter a scale per division to view specific areas
of the trace.
——Ref Level
Manually set the value of the reference line. Enter a
negative value by pressing Run/Hold (+/-)
either before or
after typing a value.
——Ref Position
Manually set the position of the reference line.
Values must be between 0 (TOP line) and 10 (BOTTOM line)
Magnitude Offset
Magnitude Offset allows you to offset the magnitude (not phase) data by a
fixed and/ or sloped value in dB. If the display format is Linear Magnitude or
Real (unitless), the conversion from dB is performed and the correct amount of
offset is implemented.
The Magnitude offset setting affects only the active trace.
How to set Magnitude Offset
—Press Scale / Amptd
.
—Then More
.
—Then Magnitude Offset
- Offsets the entire data trace by the specified
value.
Scale annotation on the FieldFox screen
Reference Line = red arrow
Ref Level = –40 dB
Ref Position = 1
Scale = 2 dB per division