Keysight N9912-90001 User’s Guide 3- 11
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
Maximum The highest level of Interference rejection.
Coupled Frequency
This setting, especially useful for a Return Loss & DTF measurement, allows
both measurements to have different frequency ranges.
How to set Coupled Frequency
With a Return Loss & DTF measurement present:
—Press Meas Setup 4
.
Select the DTF measurement (Tr2) using the arrows.
Then choose from the following:
Coupled Freq ON
- Both the Return Loss and DTF traces have the same
frequency range settings.
Coupled Freq OFF
(default setting) - Both traces are allowed to have
individual frequency range settings. When set to OFF:
The Return Loss measurement frequency settings are made in the
usual manner. Learn how at “Frequency Range” on page 3-5. When
a new Start or Stop frequency is selected, Coupled Frequency is
automatically set to OFF.
The DTF measurement is made using the frequencies as determined
by the DTF Frequency Mode setting. Learn more in “DTF
Measurement Settings” on page 4-19.