Keysight N9912-90001 User’s Guide 10- 3
Channel Power Meter Mode - Option 311
CPM Settings
CPM Settings
Connect the power to be measured to the RF IN port on the FieldFox top panel.
Select CPM Mode before making any setting in this section.
How to select Reflection Mode
—Press Mode.
—Then CPM
.
Frequency/Span/Step
Valid CPM measurements can be made over the entire frequency range of the
FieldFox. Although it is possible to enter a center frequency below the FieldFox
minimum frequency, the LO feed-thru of the internal SA hardware will
invalidate these Built-in Power Meter measurements.
The measurement span (also known as Channel Width and Integration BW) is
the frequency range, half above and half below the center frequency, over
which power is detected and displayed. The maximum measurement span is
100 MHz.
A Radio Standard can also be used to set the center frequency and span. Learn
more about “Radio Standard” on page 8-6.
How to enter center frequency, step size, and span
The default frequency is 50 MHz.
—Press Freq/Dist
or Meas Setup 4.
—Then Frequency
.
—Then Center
.
— Then type a value using the numeric keypad. Then press a multiplier key.
Learn about “Multiplier Abbreviations” on page 2-14. The current
frequency (f) is shown on the FieldFox screen.
— Optionally, use the arrows to quickly change Center frequency. The
size of the frequency step can also be changed. Press CF Step
then type a
step size value using the numeric keypad, then select a multiplier key.
Optionally press.
Please observe the RF IN port damage level.
+27 dbm (320 mW) average
+33 dBm peak (2 W) < 10μs
+50 V