8- 40 Keysight N9912-90001 User’s Guide
SA (Spectrum Analyzer) Mode
Channel Measurements
8-
By default, the displayed frequency span is automatically coupled to the
Integration Bandwidth. As you change the Integration Bandwidth, the
frequency span is adjusted so that the vertical posts appear to NOT move.
However, when you manually change the frequency span, the Integration
Bandwidth is no longer coupled to the frequency span.
When a Radio Standard is selected, the appropriate Integration Bandwidth is
set automatically. Learn more about “Radio Standard” on page 8-6.
To change Integration Bandwidth:
—Press Meas Setup 4
.
—Then Integrating BW
.
— Enter a value between 100 Hz and 3 GHz using the numeric keypad,
arrows, or the rotary knob.
—Press Enter
.
Other Channel Power Settings
All relevant FieldFox settings are made automatically to ensure the highest
accuracy, such as ResBW, VideoBW, and sweep (SwpAcquisition) speed.
These, and all other SA Mode settings, can be changed manually in a Channel
Power measurement.
—Learn about “Measurement Preset” on page 8-37.
—Learn about “Averaging” on page 8-37.
—Learn about “RRC Weighting” on page 8-46.
Occupied Bandwidth
Occupied Bandwidth measures the power of the displayed frequency span and
displays vertical posts at the frequencies between which the specified
percentage of the power is contained. The frequency span between the two
vertical posts is the Occupied Bandwidth. The Occupied Power, the power that
is contained between the two posts, is also displayed in dBm.