13-10 Keysight N9912-90001 User’s Guide
Pulse Measurements Mode – Option 330
Pulse Measurement Settings
13-
—Then Auto Trigger.
Then choose from the following:
ON
Trigger level is determined by the USB Power Sensor firmware.
OFF
The trigger level is set manually using the Trigger Level setting.
Trigger Level
Used when Trigger Type = Internal AND Auto Trigger = OFF (Manual).
—Press Sweep 3
.
—Then Trigger
.
—Then Trig Level
.
Then set the trigger level using the numeric keypad, the arrows, or the
rotary knob. A line appears on Tr 1 at the current trigger level setting. Data
acquisition occurs when the RF input signal crosses the line. This setting
can be used with either Positive or Negative edge triggering.
Trigger Settings Annotation
Trigger settings are annotated on the FieldFox screen as highlighted in red in
the following image:
In the previous graphic, the settings are: Trig Level = -10 dBm, Trig Type (Int),
Pos Slope (/), Delay