EasyManua.ls Logo

Keysight N9912A - Page 95

Keysight N9912A
344 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight N9912-90001 User’s Guide 6- 11
Time Domain – Option 010
Gating
Notch REMOVES the responses within the Gating Start and Stop times.
Gating Shape
This setting defines the filter characteristics of the gate function. Choose from
Minimum, Normal, Wide, Maximum.
Time domain Gate Shape setting is shown in the graphic above.
Cutoff time is the time between the stop time (-6 dB on the filter skirt) and the
peak of the first sidelobe.
—T
1
is the gate span, which is equal to the stop time minus the start time.
—T
2
is the time between the edge of the passband and the 6 dB point,
representing the cutoff rate of the filter.
—T
3
is the time between the 6 dB point and the edge of the gate stopband.
For all filter shapes T
2
is equal to T
3
, and the filter is the same on both sides of
the center time.
Minimum gate span is twice the cutoff time. Each gate shape has a minimum
recommended gate span for proper operation. This is a consequence of the
finite cutoff rate of the gate. If you specify a gate span that is smaller than the
minimum span, the response will show the following effects:
distorted gate shape that has no passband
—distorted shape
incorrect indications of start and stop times
may have increased sidelobe levels.
Gate Shape Passband
Ripple
Sidelobe
Levels
Cutoff Time Minimum Gate
Span
Minimum ±0.1 dB -48 dB 1.4/Freq Span 2.8/Freq Span
Normal ±0.1 dB -68 dB 2.8/Freq Span 5.6/Freq Span
Wide ±0.1 dB -57 dB 4.4/Freq Span 8.8/Freq Span
Maximum ±0.01 dB -70 dB 12.7/Freq Span 25.4/Freq Span

Table of Contents

Other manuals for Keysight N9912A

Related product manuals