6600 Product manual
Operator Manual
78
shown as error code levels and error code list.
Error Code
Error message
3
Generator CPU EEPROM Data Checksum Error
4
Generator CPU Real Time Clock error
5
Main Contactor error
6
Rotor Fault
7
Filament Fault
9
Beam_Cathode Fault
10
Beam_Anode Fault
11
Beam_INVA Fault
12
Beam_INVB Fault
13
Beam_KV Fault
14
Beam_IR Fault
15
Beam KV too low.
16
Bean kv unbalance
17
Inverter is too hot
18
Preparation Time-out Error
20
No kV during exposure
21
mA during exposure too high
22
mA during exposure too low
23
Manually Terminated Exposure
24
AEC Back-up Timer - Exposure Terminated
25
AEC MAS Exceeded - Exposure Terminated
27
Anode Heat Limit
28
Thermal Switch Interlock Error
29
Door Interlock Error
31
Bucky 1 Not Contact Error
33
Bucky 2 Not Contact Error
34
Prep Input active during Initialization Phase
35
X-ray Input active during Initialization Phase
36
Communication Error Console
37
+12VDC Error
38
-12VDC Error
43
High Voltage Error - KV detected in non x-ray state
44
Invalid Communication Message
45
Communication Message Not Supported
46
Communication Message Not Allowed
48
Current reception is not enabled
49
AEC channel is not enable in current reception
51
AEC Feedback Error (No Feedback Signal Detected)
52
High Small Focus Filament Current Error in Standby