LS SERIES SERVICE MANUAL
ISSUE 4.0 PROGRAM INFORMATION 4
Hand Held Remote
Note that there is NO diagnostic mode available if a Hand Held Remote is fitted so all
configuration and calibration MUST be performed using the NEXYGEN Plus 3
Software Console.
a Connect the NEXYGEN Plus 3 console to the machine.
b Right mouse click on the software console then select "Engineer"
c The "Engineer Passcode" screen is displayed.
d Enter 98410 as the password then click on the OK button to display the
"Engineer Options" screen
.