LogTag TRIX-8/TRIX-16 Work Instruction Manual Revision 9.20 Page 24/53
2.3 Add sealant to sensor area
Sealant options: Dow Corning 748 or 737 Non corrosive sealant.
Non-acid cure RTV is required.
Procedure:
1. Apply small amount of sealant to top of wall (& either side) where sensor passes through wall.
2. Immediately proceed to next step
Visual Inspection
Note that sealant flows into wall slots that sensor legs sit in.
Sealant must be present on both sides of the wall when case is clipped together!
Note that sealant flows into
Wall slots that sensor legs sit in.
Sealant must be present on both
sides of the wall when case is
clipped together!