LogTag TRIX-8/TRIX-16 Work Instruction Manual Revision 9.20 Page 34/53
2.11 Temperature Cycle testing
Reason for test stage:
To subject LogTag products to temperature cycle stress.
Equipment required:-
• Chest Freezer capable of achieving temperatures of -15ºC or lower.
• rack or similar to hold LogTags in place with air space between each LogTag to allow flow of air for
cooling & warming. (see appendix A.8.11)
Setup:-
• Allow Freezer to reach stable temperature for 6 hours after switch on.
Procedure
Place logtags into freezer for 12 hours
Take logtags out of freezer & into room temp for 8 hours
Place logtags into freezer for 12 hours
Take logtags out of freezer & into room temp for 8 hours
Place logtags into freezer for 12 hours
Take logtags out of freezer & into room temp for 8 hours