67
o p e r a t i n g t h e a n a l y z e r
r
SAMPLER SWITCHES
Select SAMPLER SWITCHES to verify the sampler positioning
sensor and the sample mode switch.
CAPILLARY MODE? is displayed. Lift the sampler to the capil-
lary mode and press YES.
SAMPLER CLOSED? is displayed. Push the sampler down into
the closed position.
SYRINGE MODE? is displayed. Lift the sampler to the syringe
mode and press YES.
RETURN SAMPLER is displayed. Push the sampler down
into the closed position. POSITION PASS or FAIL and
MODE PASS or FAIL are displayed. When the SAMPLER
SWITCHES test fails, refer to SENSOR MODULE in Troubleshooting.