EasyManuals Logo

Megger TM1700 Series User Manual

Megger TM1700 Series
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
52 TM1700-SERIES ZP-BL16E BL1448GE
4 CABA LOCAL
Defining pressure transducers
1]
Press the “Pressure” button on the “DEFINE
NEW TRANSDUCER” screen.
Calibrating pressure transducers can be done in two
different ways:
A] Manual. Enter all values (pairs of pressure &
voltage or current value) manually.
B] Using the instrument readings as reference.
Enter pressure value manually while using au-
tomatic TM 1800 voltage or current reading
A. Manually calibrating a pressure
transducer
1] Select the appropriate transducer output
type and if needed the power supply by
pressing the “Output/Supply” button repeat-
edly.
2] Press the “Calibrate” button. The following
question appears.
3] Press the “NO” button. The following screen
appears.
4] Enter the calibration values in the fields
above the graphic. For quick navigation be-
tween the fields use the arrow buttons next
to it.
5] Type in a significant name for the transducer
into the “Transducer ID” field.
6] Press the “Finish Calibration” button.
7] Press the “BACK” button to leave this screen.
B. TM1700 as reference for calibrating a
pressure transducer
1] Press the “Calibrate” button. The following
question appears.
2] Press the “YES” button. The following screen
appears.
3] Press the “CANCEL” button if those settings
need to be corrected and start at point 1
again. Press the “OK” button if the settings
are correct. The following message appears.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Megger TM1700 Series and is the answer not in the manual?

Megger TM1700 Series Specifications

General IconGeneral
BrandMegger
ModelTM1700 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals