MI 3152(H) EurotestXC (2,5 kV) Technical specifications
178
Test lead compensation .......................... up to 5
12.5 RCD testing
General data
Nominal residual current (A,AC) ............. 10 mA, 30 mA, 100 mA, 300 mA, 500 mA,
1000 mA
Nominal residual current accuracy .......... -0 / +0.1I; I = IN, 2IN, 5IN
-0.1I / +0; I = 0.5IN
AS/NZS 3017 selected: ± 5 %
Test current shape .................................. Sine-wave (AC), pulsed (A, F), smooth DC (B, B+)
DC offset for pulsed test current ............. 6 mA (typical)
RCD type ................................................ (non-delayed), S (time-delayed), PRCD, PRCD-K,
PRCD-S
Test current starting polarity .................. 0º or 180º
Voltage range ......................................... 93 V ... 134 V (45 Hz ... 65 Hz)
185 V ... 266 V (45 Hz ... 65 Hz)
RCD test current in relation to RCD type, nominal RCD current and multiplication factor
× ............................................................. not applicable
............................................................. applicable
AC type................................................... sine wave test current
A, F types…… ........................................ pulsed current
B, B+ types ............................................. smooth DC current (MI 3152 only)
RCD test current in relation to MI / EV RCD type and multiplication factor
×. ............................................................ not applicable
. ............................................................ applicable
MI / EV types (a.c. part) .......................... sine-wave test current
MI / EV types (d.c. part) .......................... smooth DC current
12.5.1 RCD Uc – Contact voltage
Measuring range according to EN 61557 is 20.0 V ... 31.0 V for limit contact voltage 25 V
Measuring range according to EN 61557 is 20.0 V ... 62.0 V for limit contact voltage 50 V