MI 3152(H) EurotestXC (2,5 kV) Technical specifications
181
12.8 Zs rcd – Fault loop impedance and prospective fault
current in system with RCD
Z Fault loop impedance
Measuring range according to EN 61557 is 0.46 ... 9.99 k for I test = standard and 0.48 ...
9.99 k for I test = low.
Measuring range ()
Resolution ()
Accuracy
I test = standard
Accuracy
I test = low
0.00 ... 9.99
0.01
(5 % of reading + 10
digits)
(5 % of reading + 12
digits)
10.0 ... 99.9
0.1
100 ... 999
1
10 % of reading
10 % of reading
1.00 k ... 9.99 k
10
Accuracy may be impaired in case of heavy noise on mains voltage.
Ipsc Prospective fault current
Measuring range (A)
Resolution (A)
Accuracy
0.00 ... 9.99
0.01
Consider accuracy of fault
loop resistance measurement
10.0 ... 99.9
0.1
100 ... 999
1
1.00 k ... 9.99 k
10
10.0 k ... 23.0 k
100
Ulpe Voltage
Measuring range (V)
Resolution (V)
Accuracy
0 ... 550
1
(2 % of reading + 2 digits)
Uc Contact voltage
Refer to chapter 12.5.1 RCD Uc Contact voltage for detailed technical specification.
Nominal voltage range ........................... 93 V ... 134 V (45 Hz ... 65 Hz)
185 V ... 266 V (45 Hz ... 65 Hz)
No trip out of RCD. R, X
L
values are indicative.
12.9 Z loop m – High precision fault loop impedance and
prospective fault current
Refer to A 1143 Euro Z 290 A adapter Instruction manual for detailed technical
specification.