MI 3152(H) EurotestXC (2,5 kV) Tests and measurements
157
7.29 Z auto - Auto test for fast line and loop testing
Tests / measurements implemented in Z auto test sequence
Figure 7.96: Z auto menu
Measurement parameters / limits
Protection type [TN, TNrcd, TTrcd]
Selection of fuse type [Off, gG, NV, B, C, D, K, Custom]
Rated current of selected fuse
Maximum breaking time of selected fuse
Rated current for U measurement (custom value)
Isc factor [0.20 ... 3.00]
Test current [Standard, Low]
RCD type [AC, A, F, B*, B+*,F]
Rated RCD residual current sensitivity [10 mA, 30 mA, 100 mA, 300 mA,
500 mA, 1000 mA]
Selection of test [-, L1, L2, L3]
Test current [Standard, Low]
Maximum voltage drop [3.0 % ... 9.0 %]
Ia(Ipsc (LN),
Ipsc (LPE))
3)
Minimum short circuit current for selected fuse or custom value
Conventional touch voltage limit [12 V, 25 V, 50 V]
1)
Applicable if Fuse type is set to Off or Custom.
2)
With Plug test cable or Plug commander Z auto test is measured in the same way
regardless of the setting. The parameter is meant for documentation.
3)
Ipsc (LPE) is considered if Protection is set to TN or TNrcd. Ipsc(LN) is always
considered.
* Model MI 3152 only
Refer to Fuse tables guide for detailed information on fuse data.