401- The communications port specified for the SmartPrep(s) is already in
use. Cannot initialize.
Cause: The communications port associated with this instrument is in
use by some other program in the system.
Action: Close the other program to release the port. Restart the
SmartPrep 065 program.
402- The communications port specified for the SmartPrep(s) cannot be
accessed. Cannot initialize.
403- Cannot communicate with SmartPrep Unit %d - S/N: %s.
Cause: The communications port associated with this unit was not valid.
Action: Run the setup program and set up the unit on a valid port.
404- The application version of the SmartPrep Unit %d - S/N: %s is invalid.
Cause: The controller software running on the designated instrument is
invalid.
Action: Use the SmartPrep Setup program to download the proper con-
troller software to the instrument, or contact a service repre-
sentative if the Setup program is unavailable.
405- Fatal communications error with SmartPrep Unit %d - S/N: %s.
Cause: There was a fatal error in the serial communications between
the application and the SmartPrep Instrument Controller. All dis-
plays for that SmartPrep will be closed.
Action: Check that the SmartPrep is properly connected to the control
module on the communications port configured in the Setup pro-
gram. Stop and re-start the application. If this error persists,
contact your service representative.