Status Indicators Each heating station has three lights that indicate
the current status of the station. The lights flash
when gas is flowing through the heating station.
Ready The green Ready light comes on when the set-
point temperature is below 30
o
C and the actual
temperature is below 60
o
C (the station is idle
and is not hot).
Cooling The yellow Cooling light comes on when the set-
point temperature of the station is less than 30
o
C
and when the actual temperature is greater than
60
o
C.
Heating The red Heating light comes on when the setpoint
temperature is greater than 30
o
C and heat is be-
ing applied to the station.
Front Panel Controls and Indicators SmartPrep
2-2 Aug 98