EasyManua.ls Logo

Milestone LOGOS One - Page 85

Milestone LOGOS One
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Milestone s.r.l.
MM098-002 LOGOS ONE Operator Manual
85
Dispose of the exhausted reagent.
Load the tank with the FRESH REAGENT (Flushing MIX in this case).
Put the tank containing fresh reagent into the position indicated by the software (number 2 in this case).
Select the reagent in the drop-down menu. The software, as a default, suggests filling with the same reagent
already loaded into the tank. To load a different reagent, select it from the drop-down menu (figure below,
circle). Press Done to confirm and return directly to the REAGENT CONFIGURATION screen.
Close the tanks holder doors and check the reagent level (working label).
Settings are automatically saved by the software.
After the replacement, the software resets the reagent management count to zero and the loading date
automatically changes to the current date (figure below). The reagents status icon also turns from red to
GREEN (figure below).
The software always suggests the last reagent loaded.
Make sure tank fitting is correctly inserted in its relative housing, and that the tank is
aligned with the others.

Table of Contents

Related product manuals