116
2. Clean by spraying the concerned surface and pass with a cloth.
3. For decontamination, start the automatic procedure with VIROCLEAN.
4. Allow it to act for at least 5 to 30 minutes (refer to the previous table for the contact times) depending on
the level you wish to reach according to the criticality of the treated device:
a. bactericidal and yeasticidal: 5 minutes;
b. fungicidal and virus inactivating: 15 minutes;
c. tuberculocidal and sporicidal: 30 minutes;
4.3. Operating procedure with 70% alcohol-based reagent
1. Clean by spraying the concerned surface and pass with a cloth.
2. For decontamination, spray the concerned surface.
3. Leave the reagent to act according to that set forth in the technical data sheet.
4. If necessary, rinse and dry according to that set forth in the reagent technical data sheet.
4.4. Preventive maintenance
The software on the UltraGROSS terminal notifies the operator when maintenance needs to be done (default
1 year) and when the filters (if any) need to be changed, as explained in chapter 2.6.
Annually contact your local dealer or Milestone offices directly:
customersupport@milestonemedsrl.com for how to perform maintenance and regular
inspections in order to maintain UltraGROSS within safe operative standards.