2.5.17. Installation checks
Refer to the installation documents. If not available, please contact our Customer support at:
customersupport@milestonemedsrl.com.
Before switching on unit ventilation with active hood (letter A):
• Check that all the alarms, reported in chapter 2.6, work properly.
• Check that the safety shield is intact.
• Check that the coupling of the safety shield is in the “lock” position.
• Check that the Ø 160mm air outlet pipes are intact.
• Check that the plastic ducts are intact.
• Check the position of the anemometer.
• Check the correct fixing of the front baffle and of the extraction panel.
• Check tightening of the screws on the rear panels.
• Check that the pressure switch pipe is not crushed and is connected.
• Check that there are no air leaks along the ducts.
• Check that both motors work properly.
Switch on ventilation using the membrane keyboard (see chapter 2.6), remove whatever is not strictly
necessary on the work surface and using the anemometer/barometer, check that:
• Speed is constant along the entire length of the front opening (with the maximum deviation of 0.1m/s).
• The average speed reaches at least 0.5 m/s and does not exceed 1m/s (thresholds set in General settings
- see chapter 2.6).
Before switching on unit ventilation with passive hood (letter P):
• Check that all the alarms work properly.
• Check that the safety shield is intact.
• Check that the coupling of the safety shield is in the “lock” position.
• Check that the Ø 160mm air outlet pipes are intact.
• Check that the two “passive hood” plastic ducts are intact.
• Check the position of the anemometer.
• Check the correct fixing of the front baffle and of the extraction panel.
• Check tightening of the screws on the rear panels.
• Check that the pressure switch pipe is not crushed and is connected.
• Check that there are no air leaks along the ducts.
Switch on ventilation of the external motor, remove whatever is not strictly necessary on the work surface and
using the anemometer/barometer, check that:
• Speed is constant along the entire length of the front opening (with the maximum deviation of 0.1m/s).
• The average speed reaches at least 0.5m/s and does not exceed 1.1m/s (thresholds set in Factory
settings - see chapter 2.6)
If one of the above indicated points is not confirmed, check the fume extraction system outside of the unit.
Lastly, UltraGROSS must be placed in the final position for use.
Unscrew the levelling feet to lift the five casters off the ground. Then make sure that the unit is level on a
horizontal surface.
The installation part has finished successfully.