UltraGROSS
MM182-001 Operator Manual
115
4.2.1. MANUAL operating procedure
The product is ready for use and therefore requires no dilution.
1. Insert the supplied dispenser.
2. Clean by spraying the concerned surface and pass with a cloth.
3. For decontamination, spray the concerned surface.
4. Allow it to act for at least 5 to 30 minutes (refer to the previous table for the contact times) depending on
the level you wish to reach according to the criticality of the treated device:
a. bactericidal and yeasticidal: 5 minutes;
b. fungicidal and virus inactivating: 15 minutes;
c. tuberculocidal and sporicidal: 30 minutes;
5. Dry if necessary.
4.2.2. AUTOMATIC operating procedure
The product is ready for use and therefore requires no dilution.
Cleaning must always be done in manual mode. Decontamination can be done in automatic
mode with the VIROCLEAN instrument present (chapter 5.31).
Automatic decontamination is limited solely to the area enclosed by the protective shield,
as shown in the following image.
1. Insert the supplied dispenser.
Switch off UltraGROSS and disconnect the power cable.