UltraGROSS
148
5.31. VIROCLEAN
CODE 109351A
The unit works with a dry hydrogen peroxide mist. Use a heating and ionizing turbine that sprays the
disinfectant liquid after it is transformed into an extremely dry mist. This mist is non-wetting and without
moisture, with an average particle size smaller than 5µm. It guarantees a quick and even saturation on the
treated surfaces.
The unit is supplied without liquid; refer to the next chapter.
Before starting the process, the surfaces must be cleaned.
Packaging size: WxDxH 60 x 44 x 42cm.
Instrument size: WxDxH 45 x 44 x 42cm.
Weight: 10Kg.
Absorption: 1000W.
Operating voltage: 230V~50Hz.
The power cable must be connected to a socket outside of the unit.
For further information on the instrument, see the manual supplied with UltraGROSS.
VIROCLEAN includes: instrument, pipe with dispenser, power cable.
INSTALLATION AND OPERATION:
- the instrument can be placed on the ground either on the right or left of the unit.
- Remove the closing cap on the glass.
- Insert the dispenser into the hole on the glass and block it with the specific ring nut.
- Connect the reagent bottle in position without crushing the dip tube.