19
1.3.3. Technical specifications
For letters that refer to unit codes, see chapter 1.3.1.
For voltage and frequency in reference to letters E customizable like K (220-240V~ 50/60Hz) the power is:
• 2600W if letter (A) active 1819BTU/h,
• 1070W if letter (P) passive 512BTU/h.
For voltage and frequency in reference to letter U (115V~ 60Hz) the power is:
• 2160W if letter (A) active 1836BTU/h,
• 610W if letter (P) passive 512BTU/h.
For voltage and frequency in reference to letter J (100V~ 50/60Hz) the power is:
• 1500W if letter (A) active 1666BTU/h,
• 150W f letter (P) passive 512BTU/h.
For codes that start with 180:
o If letter A: Weight:350Kg – 771.61lb,
o If letters A and C: weight 375Kg – 826.73lb,
o If letter P: weight :325Kg – 716.50lb,
o If letters P and C: weight 350Kg – 771.61lb,
o For VIRCHOW model, if letter A, weight 400Kg – 881.84lb,
o For VIRCHOW model, if letter P, weight 375Kg – 826.73lb.
For codes that start with 150:
o If letter A: Weight:325Kg – 716.50lb,
o If letters A and C: weight 350Kg – 771.61lb,
o If letter P: weight :300Kg – 661.38lb,
o If letters P and C: weight 325Kg – 716.50lb.
For the sizes of all models, see chapters 2.2 and 2.3.
Lighting
- LED illumination of the work area
- Average work area illuminance 1600 lux*
* with environmental conditions of 400 Lux and units with all lights on
- Uniformity U=0,9
- Color rendering index Ra=70
Structure
- Work surfaces in AISI 316 stainless steel (listed in chapter 5.1).
- Structure in AISI 304 stainless steel
- Antibacterial paint
-Work surface height adjustable from 85cm to 115cm
- Telescopic columns (Protection rating IPx6)
- Safety shield in shatterproof glass
- Tap with clinical lever
- Shelf
- Magnetic holder bar
- The instrument has 6 casters
- Object shelf outside of work surface
Types of seat (EN 14738:2009)
Compatible with machine ergonomics:
- Lifted sitting seat.
- Upright posture with support.
- Upright posture.