2.5.9. Connection to the hydraulic system Water drain
Wastewater connection (back of the unit):
Connect the 5m long flexible pipe Ø40mm to the unit drain, adjusting its length and tighten the strap.
After mounting the pipe, insert it into the relevant support hook.
Connect the opposite end to the laboratory's drain.
The basin underneath the work surfaces is connected to the drain.
The drain system is not equipped with siphon.
The laboratory's wastewater drain must be lower than 40cm to ensure optimal drainage. If
a Waste disposal unit is used, see chapter 5.10.
Make sure that the drain pipe has a slope of approximately 10cm per meter (unit at minimum
height).
Do not plug the basin drain.
Do not use the taps if the drain is clogged. Contact Customer support.