MAINTENANCE
Milli-Q Synthesis/Synthesis A10 68
It is very important that the POU Dispenser is closed now. The Sanitisation
timer will not count down from 400 minutes until the valve is kept closed.
When the timer reaches 1 mn, the Milli-Q will automatically go into PRE OPERATE Mode.
The Sanitisation and Cleaning of the UF Cartridge is now finished.
The water inside the Milli-Q will be recirculated for 5 minutes every hour of PRE OPERATE Mode.
When you use the system (i.e. the morning after the UF was Sanitised), dispense enough water until the Resistivity
(and TOC) values are satisfactory.
Remember to leave the Milli-Q in PRE OPERATE Mode during times of non use (i.e. evenings,
weekends, holidays,...).