4 - 40 Operator’s Manual
4 Setup
1. Select the target module.
2. Click [Export].
3. Select the path to save the data.
4. Select the exported file and type as PDP and click [OK].
4.10.3 Importing Setup Data
This function is used to import the existing setup data to the setup data memory of the system. The
system will reset and operate according to the setup preferences that were imported.
Perform the following procedure:
1. Click [Import].
2. Select the imported file.
3. Click [OK], a progress bar will appear and the setup data is imported to the specified path.
4.10.4 Load Factory
To restore the factory setup data, click [Load Factory] on the right side of the screen.
4.10.5 Probe Check
This function enables users to check if a transducer element is in malfunction, so as to evaluate the
transducer performance.
Before performing the probe check, ensure the probe head must be clean and the probe is in
nonscanning state.
Click [Probe Check] to open the Probe Check screen. The system automatically performs probe
check to the element of the currently activated probe.
• If a transducer element is in malfunction, it is displayed as a red spot.
• If a transducer element functions well, it is displayed as a green spot.
You can export the result image to the external device.
4.10.6 Other Settings
Other preset settings are described as follows:
If you have any questions, please contact the service engineer or your sales representative.
Item Description
Setup Export Log Export the log.
Self Test Perform system self-test and restart the machine.
Recover To recover the system.
Battery Manufacture
State
Querying Battery Health Status.
Prepay Installment Display the prepay installment information.