4 - 4 Operator’s Manual
4 Setup
4.1.3 Image Preset
Keyboard
Sound
Keyboard sound
enabled
Turn on/off the keyboard sound.
Type Item Description
Type Item Description
Reset Config Probe To set the default probe model for the system from the drop-
down list.
The default parameters are applied to the new probe if
checking “Use the default setting when start a new exam.”
Image Size Set the standard for saving the image or using the digital/
graph printer.
Default Elasto Set the default elastography mode.
Parameter Steer To set the steer mode in B + Color + PW/CW imaging mode.
• C&(PW/CW): select to adjust the sample volume in color
mode and sample line in PW or CW mode together.
• C/(PW/CW): select to adjust the sample volume in color
mode and sample line in PW or CW mode separately.
Auto Invert The spectrum can automatically invert when the color flow
is steered to a certain angle, thus accommodating the
operator’s wish to distinguish the flow direction.
iScape Ruler
Display
To set whether to display the iScape ruler in iScape imaging
mode.
B+Color Refresh
With PW/CW
Sampling Line
Movement
To set whether to turn on the function that when moving
PW/CW sampling line, B+Color image is activated under
B+Color+PW/CW mode.
Uninterruptible
image saving when
preprocessing
parameters are
changed
Adjusting preprocessing parameters will not interrupt image/
cine saving.
Color And PW/CW
Synchronize Invert
To set whether to invert Color Map and PW/CW spectrum
synchronously
Display Transducer
SN
To set whether to display the transducer SN.
B+Color+PW
Unfreeze keeping
PW
To set whether to retain PW images when unfrozen under
B+Color+PW mode.
iTouch+ • If selected, the system optimizes the images continuously
in real time after you press < iTouch
+
> in B mode.
• If not selected, the system optimizes the images only
once after you press < iTouch
+
> in B mode.