B - 2 Operator’s Manual
B Wireless LAN
SSID > 80% with stable WLAN network.
Wireless router and the server are in the same network segment.
Router setting:
Wireless standard IEEE 802.11 ac/a/b/g/n.
Maximum transmission speed 300 Mbps.
Number of the devices connected to the same router 5.
Target server setting:
Network is stable and not under overloading state (e.g. high CPU/memory usage, fast
HDD speed, limited HDD space).
Level other than the highest level of firewall is adopted.
Operating system is Windows 10 or higher versions and it supports a Gigabit Ethernet.
B.1 Use the Wireless Feature
Perform the following procedure:
1. Press <Cursor> to show the cursor, click in the bottom bar to open the wireless network
manager.
2. Move the cursor to the target network and press <Set> to select it, then click [Connect] to
connect to the network.
When connecting an encrypted network, enter the password in the box first. You can select to
hide password characters or not.
3. The system tries to connect and the wireless manager icon turns into . The icon turns into
or after successful connection.
4. Click [Refresh] to refresh the “Wireless Network Connection” list.
B.2 IP Configure
NOTE:
When the system background is processing network task (DICOM sending for example),
please do not enter network setting to change the IP, otherwise the background task may fail.
You can check if there are tasks undergoing in the task manager.
If the IP address displays as 0.0.0.0, this means that the network is abnormal. The reason for
the failure may be disconnection or the system cannot obtain the IP address.
IP Config is used for setting local network parameters, which is also applied to DICOM connection.
Perform the following procedure:
1. In Wireless network manager screen, click [IP Config] to open the configuration page.
If “DHCP” is selected, the IP address will be automatically obtained from the DNS server.
If “Static” is selected (using a static IP address), enter the IP address.
IP address of the system should be in the same network segment with the server.
The name of the device is saved under the service name by default. The system
remembers the service name of the ultrasound system when sending the image, the report
to DICOM server. Open the file (DCM Editor Tool, eZDicom.exe) to view the service
name (iStation Name).