EasyManua.ls Logo

Mindray Consona N6 - 2.4 Product Specifications

Mindray Consona N6
396 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 - 2 Operator’s Manual
2 System Overview
Does the equipment has any defibrillation-proof applied parts:
ECG is equipped with the applied part of defibrillation protection.
Permanently installed equipment or non-permanently installed equipment:
Non-permanently installed equipment
2.4 Product Specifications
NOTE:
The functions described in the operators manual may vary depending on the specific system
purchased.
2.4.1 Power supply
Power input
Voltage: 100 - 240 VAC
Frequency: 50/60 Hz
Power input: 6.2-2.0A
Battery voltage: 14.4 VDC, 6600 mAh
Power output
Voltage: 100 - 240 VAC
Frequency: 50/60 Hz
Power consumption: 240VA (This is the maximum auxiliary output power of the outlet. When
the peripheral device is connected to the outlet, ensure that the maximum auxiliary output
power does not exceed this threshold. Otherwise, the system may be damaged.)
2.4.2 Environmental Conditions
WARNING
Do not use this system in conditions other than those specified.
Operating conditions
Ambient temperature: 0 °C ~ 40 °C
Relative humidity: 20% ~ 85% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
Storage and transportation conditions
Ambient temperature: -20 °C ~ 55 °C
Relative humidity: 20% ~ 95% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
2.4.3 Dimensions and Weight
Dimensions (The monitor and the control panel are kept to a minimum position):
Length: 840±40mm, Width: 513±10mm, Height: 994±20mm
Weight: 65.0 kg (Standard configuration)

Table of Contents