19 - 2 Operator’s Manual
19 System Maintenance
2. Use soft brush to brush the dust inside probe connector gently.
3. Remained stain or dust attached to surface of cable or surface of connector should be washed
out by cloth with a little soapy water, and then air-dry.
Don’t use cloth with water to clean the probe connector.
Cleaning holders
Clean the holders periodically (1 time per month).
Tools: soft dry cloth, soapy water, soft brush.
Remaining stains should be wiped away using a cloth with clean or soapy water and the surface left
to air dry.
Perform the following procedure:
1. Use a soft dry cloth to wipe away dust attached to the inside, outside and gaps in the probe
holder. Use a soft brush to brush away dust or stains from the small intra-cavity probe holder or
its gap.
2. Remaining stains on the inside and outside of the holder should be wiped away using a cloth
with a little soapy water and then air dried.
Cleaning the machine shell
• Use a soft brush to gently remove the dust from naked interfaces or sockets (such as probe
sockets, IO panels, and power supply panel). Do not use a water cloth.
• Clean the machine shell periodically (1 time per month).
Tools: mild soapsuds, and dry soft cloth
Use a dry soft cloth to wipe the dirt off the machine shell (the exposed part).
Or, use with a dry soft cloth dipped in a small amount of mild soapsuds to remove stains, and air dry
the shell.
Cleaning the monitor and the touch screen
• DO NOT use hydrocarbon glass cleaner or cleaner for OA (Office Automation) equipment to
clean the monitor. These substances may cause deterioration of the monitor.
• Clean the monitor and the touch screen periodically (1 time per month).
Tool: soft dry cloth, soapy water
The surfaces of the monitor and touch screen should be cleaned with a soft dry cloth. Remaining
stains should be wiped away using a cloth with a little soapy water and then air dried.