A-13
ELECTRICAL SAFETY INSPECTION FORM
Location: Technician:
Equipment: Control Number:
Manufacturer: Model: SN:
Measurement equipment /SN: Date of Calibration:
INSPECTION AND TESTING Pass/Fail Limit
1 Power Cord Plug
2 Device Enclosure and Accessories
3 Device Labeling
4 Protective Earth Resistance ____ Max 0.2
Normal
condition(NC)
____μA
5
Earth
Leakage
Single Fault
condition(SFC)
____μA
Max:
NC: 300μA(refer to
UL60601-1) *
NC: 500μA(refer to
IEC60601-1) *
SFC: 1000μA
BF____μA
Normal
condition(NC)
CF____μA
BF____μA
6
Patient
Leakage
Current
Single Fault
condition(SFC)
CF____μA
Max:
CF applied part:
NC:10μA, SFC: 50μA
BF applied part:
NC:100μA, SFC:
500μA
BF____μA
7 Mains on Applied Part Leakage
CF____μA
Max:
CF applied part: 50μA
BF applied part:
5000μA
BF____μA
Normal condition(NC)
CF____μA
BF____μA
8
Patient
Auxiliary
Current
Single Fault
condition(SFC)
CF____μA
Max:
CF applied part:
NC:10μA, SFC: 50μA
BF applied part:
NC:100μA, SFC:
500μA
Note: The equipment sold to the United States shall comply with the requirement of
UL60601-1; others shall comply with the requirement of IEC60601-1.
Name/ Signature: __________________________ Date:_____________________________

Table of Contents

Other manuals for Mindray DPM 6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 6 and is the answer not in the manual?

Mindray DPM 6 Specifications

General IconGeneral
BrandMindray
ModelDPM 6
CategoryHeart Rate Monitor
LanguageEnglish

Related product manuals