3-17 
3.4.5 Temp Test 
Tool required: 
  Resistance box (with accuracy above 0.1Ω) 
 
Follow this procedure to perform the test: 
1.  Connect the two pins of any Temp connector of a module to the two ends of the 
resistance box using 2 wires. 
2.  Set the resistance box to 1354.9Ω (corresponding temperature is 37ºC). 
3.  Verify that the displayed value is within 37 ± 0.1ºC. 
4.  Repeat steps 1 to 3 and verify another temperature channel. 
 
 
3.4.6 IBP Tests 
IBP Performance Test 
Tool required: 
  Medsim300B patient simulator, MPS450, or other equivalent device 
  Dedicated IBP adapter cable for test (P/N 009-002199-00 for Medsim 300B, P/N 
009-002198-00 for MPS450)   
 
Follow this procedure to perform the test: 
1.  Connect the patient simulator with the pressure module. 
2.  Make the patient simulator outputs 0 to an IBP channel. 
3.  Press the Zero Key on the module to make a zero calibration. 
4.  Configure the patient simulator as P (static) = 200 mmHg. 
5.  The displayed value should be within 200 ± 4 mmHg. 
6.  If the error is beyond ±4 mmHg, calibrate the pressure module. If the IBP module was 
calibrated with a dedicated reusable IBP sensor, check the calibration together with this 
IBP sensor. 
7.  Make the patient simulator output 120/80 mmHg ART signal and 120/0 mmHg LV 
signal to the IBP channel and check that the IBP wave is displayed correctly. 
8  Repeat the steps above for all the IBP channels.