3-16 
 
3.4.2 Resp Performance Test 
Tool required: 
  Fluke Medsim 300B patient simulator recommended 
Follow this procedure to perform the test: 
1.  Connect the patient simulator to the module using a non ESU-proof cable and set lead II 
as the respiration lead. 
2.  Configure the simulator as follows: lead II as the respiration lead, base impedance line 
as 1500 Ω; delta impedance as 0.5 Ω, respiration rate as 40 rpm. 
3.  Check the Resp wave is displayed without any distortion and the displayed Resp value is 
within 40 ± 2 rpm. 
 
3.4.3 SpO
2
 Test 
Tool Required:   
  None. 
 
Follow this procedure to perform the test: 
1. Connect SpO
2
 sensor to the SpO
2 
connector of the monitor. Set [Patient Cat.] to [Adu] 
and [PR Source] to SpO
2
 on the monitor. 
2.  Apply the Measure SpO
2
 sensor to on your ring finger. (Assume that you stay healthy) 
3.  Check the Pleth wave and PR reading on the screen and make sure that the displayed 
SpO
2
 is within 95% and 100%. 
4.  Remove the SpO
2
 sensor from your finger and make sure that an alarm of SpO
2
 Sensor 
Off is triggered. 
 
NOTE 
z  A functional tester cannot be used to assess the accuracy of a pulse oximeter 
monitor. However, it can be used to demonstrate that a particular pulse oximeter 
monitor reproduces a calibration curve that has been independently demonstrated 
to fulfill a particular accuracy specification. 
 
 
3.4.4 NIBP Tests 
Refer to 3.2.2 NIBP Tests and Calibration.