3-2
3.1.2 Test Report
Upon completion of the tests, the table of preventative maintenance test reports and the table
of maintenance test reports in this chapter should be kept properly.
3.1.3 Preventative Maintenance
Below are preventative maintenance tests which need to be performed on the monitor. See
the following sections for detailed maintenance procedures.
Visual inspection
NIBP test and calibration
Microsteam and Sidestram CO
2
test and calibration
AG test and calibration
3.1.4 Recommended Frequency
Check/Maintenance Item Frequency
Preventative Maintenance Tests
Visual inspection 1. When first installed or reinstalled.
Pressure check
Leakage test
NIBP test
Calibration
Leakage test
Performance test
Sidestream and
Microstream
CO
2
tests
Calibration
Leakage test
Performance test
AG tests
Calibration
1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of relevant
module.
3.At least once a year is recommended.
Performance Tests
Performance test ECG test and
calibration
Calibration
Resp
performance test
/
1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of relevant
module.
3. At least once every two years.

Table of Contents

Other manuals for Mindray DPM 6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 6 and is the answer not in the manual?

Mindray DPM 6 Specifications

General IconGeneral
BrandMindray
ModelDPM 6
CategoryHeart Rate Monitor
LanguageEnglish

Related product manuals