Respiration and CO
2
Troubleshooting Respiration Monitoring
5 - 6 0070- 0-0704-02 Passport V Operating Instructions
Resp Disturbed Patient Movement
External Interference.
Check patient
For impedance - Check
Electrode Contacts / reposition
electrodes/cable
Check environment for source
of interference.
Resp Rate Overrange Respiration Outside of
measureable range
Check patient.
Resp: Initialization Error During the Resp module power-
on,as the Resp module
communication stops,system
fails to communicate with
module.
Contact Technical Support.
Resp: Communication Error System can't communicate
correctly with Resp module,and
receive the error data packets.
Restart the monitor.If the error
still appears,please contact
Technical Support.
Resp: Communication Stop As the Resp module
communication stops, the data
packets sent by the module
can not be received.
Contact Technical Support.
Resp: High Impedance Connections not tight and/or
properly secured.
Electrodes dry or loose.
Cable or lead wires damaged.
Ensure proper connection.
(Electrode to lead, lead to
cable, cable to monitor).
Re-prep skin and apply fresh,
moist electrodes.
Check with continuity tester
MESSAGE/PROBLEM REASON SOLUTION