IntroductionAvoid simultaneous contact with the patient and the SP1 and SP2 ports on the Passport V to avoid excessive leakage current.
xvi 0070- 0-0704-02 Passport V Operating Instructions
CAUTION: Devices connected to the Passport V must meet the
requirements of the applicable IEC standards (e.g., IEC 60950
safety standards for information technology equipment and IEC
60601-1 safety standards for medical electrical equipment). The
system configuration must meet the requirements of the IEC
60601-1-1 medical electrical systems standard. Any personnel
who connects devices to the signal input/output port of the
Passport V is responsible for providing evidence that the safety
certification of the devices has been performed in accordance to
the IEC 60601-1-1. Please contact Technical Support if you have
any questions about connecting devices to the data port.
CAUTION: The Gas Module must be power cycled if the Passport V monitor
is powered off.
Avoid simultaneous contact with the patient and the SP1 and SP2 ports on the
Passport V to avoid excessive leakage current.
Notes
NOTE: This unit is not designed to be used with a peripheral pulse
sensor. SpO
2
is a standard function in this monitor, and may be
used to obtain a plethysmograph waveform and heart rate.
NOTE: The comparison testing conducted via the auscultatory method
used both Phase 4 and Phase 5 Korotkoff sounds. Reports of
study findings for both the auscultatory method as well as the
intra-arterial methods are available by contacting Technical
Support (800) 288-2121, ext. 8116 or (201) 995-8237.
NOTE: Potential hazards due to errors in software or hardware have
been minimized by actions taken in accordance with IEC 60601-
1-4.
0