Prompt Messages Messages
14 - 8 0070- 0-0704-02 Passport V Operating Instructions
14.3 Prompt Messages
14.3.1 ECG Prompt Message
14.3.2 Resp Prompt Message
ALARM MESSAGE REASON SOLUTION
ECG Self Test Error During Power cycle unit, if ECG
module can not self test
succussfully, this message will
display.
Power cycle unit. If message
reappears, contact Technical
Support.
ECG: Communication Stop As the ECG module
communication stops, the data
packets sent by the module can
not be received.
Contact Technical Support.
ECG: Communication Error The ECG module communication
error, the command can not be
send correctly.
Contact Technical Support.
Leads Off Leads off patient. Connect lead wires.
ESU-Resp Off The high frequency
electrosurgery Unit
Interference signal is detected.
Wait that the high frequency
electrosurgery Unit
Interference disappears.
ECG Signal Invalid Poor leads connection. Check Patient.
Check Leads and leadlines.
Pacer Rejection On When Pacer Reject is set to On. This is normal operation. When
Pace Reject is set to Off, this
message disappears.
Learning Displayed when a learning cycle
has been requested for
Arrhythmia or ST.
/
No Arrhythmia Detection at
Central
Central Station does not have
arrhythmia Analysis capability.
/
ECG: Initialization Error During the ECG module power-
on,as the ECG module
communication stops,system fails
to communicate with module.
Contact Technical Support.
ALARM MESSAGE REASON SOLUTION
Resp: Initialization Error During the Resp module power-
on,as the Resp module
communication stops,system fails
to communicate with module.
Contact Technical Support.
Resp: Communication Error System can't communicate
correctly with Resp module,and
receive the error data packets.
Restart the monitor.If the error
still appears,please contact
Technical Support.
0

Table of Contents

Other manuals for Mindray Passport V

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish